Iodine quantification and detectability thresholds among major dual-energy CT platforms
To estimate the minimum detectable iodine concentration on multiple dual-energy CT (DECT) platforms. A phantom containing iodine concentrations ranging from 0 to 10 mg ml was scanned with five dual-energy platforms (two rapid kilo volt switching (r-kVs), one dual source (DS), one sequential acquisit...
Gespeichert in:
Veröffentlicht in: | British journal of radiology 2019-12, Vol.92 (1104), p.20190530-20190530 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | To estimate the minimum detectable iodine concentration on multiple dual-energy CT (DECT) platforms.
A phantom containing iodine concentrations ranging from 0 to 10 mg ml
was scanned with five dual-energy platforms (two rapid kilo volt switching (r-kVs), one dual source (DS), one sequential acquisition and one split-filter). Serial dilutions of 300 mg ml
iodinated contrast material were used to generate concentrations below 2 mg ml
. Iodine density and virtual monoenergetic images were reviewed by three radiologists to determine the minimum visually detectable iodine concentration. Contrast-to-noise ratios (CNRs) were calculated.
1 mg mL
(~0.8 mg mL
corrected) was the minimum visually detectable concentration among the platforms and could be seen by all readers on the third-generation r-kVs and DS platforms.
At low concentrations, CNR for monoenergetic images was highest on the DS platform and lowest in the sequential acquisition and split-filter platforms.
The results of this study corroborate previous
estimates of iodine detection limits at DECT and provide a comparison for the performance of different DECT platforms at low iodine concentrations
. |
---|---|