Decontamination of Intravaginal Probes Infected by Human Papillomavirus (HPV) Using UV-C Decontamination System
This three-step study evaluated ultraviolet-C (UV-C) efficacy against human papillomavirus (HPV) found on vaginal ultrasound probes. The first two steps evaluated UV-C disinfection of vaginal ultrasound probes in routine condition. During the first phase, the probe ( = 100) was sampled after a compl...
Gespeichert in:
Veröffentlicht in: | Journal of clinical medicine 2019-10, Vol.8 (11), p.1776 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | This three-step study evaluated ultraviolet-C (UV-C) efficacy against human papillomavirus (HPV) found on vaginal ultrasound probes.
The first two steps evaluated UV-C disinfection of vaginal ultrasound probes in routine condition. During the first phase, the probe (
= 100) was sampled after a complete cleaning and disinfection protocol, i.e., cleaning with chemically impregnated wipes, followed by UV-C. During the second phase, the probe (
= 47) was sampled after cleaning and UV-C. The final step consisted of applying mixes of HPV on a dedicated, covered probe (
= 15) then sampling the cover, the probe after removal of the cover, after cleaning, and after UV-C. HPV detection was performed using CLART
HPV2 PCR (Genomica, Madrid, Spain).
In the first phase, no probes were found to be positive for both DNA after UV-C. In the second phase, eight probes were found to be positive after cleaning (seven with human DNA and one with HPV) and negative after UV-C. In the final phase, one probe was found to be positive for HPV for each sample except after UV-C.
Covers followed by a chemically impregnated wipe are not sufficient to ensure patient safety during vaginal ultrasound examinations. UV-C is effective in routine conditions against contaminations found on vaginal ultrasound probes, especially HPV. |
---|---|