Decontamination of Intravaginal Probes Infected by Human Papillomavirus (HPV) Using UV-C Decontamination System

This three-step study evaluated ultraviolet-C (UV-C) efficacy against human papillomavirus (HPV) found on vaginal ultrasound probes. The first two steps evaluated UV-C disinfection of vaginal ultrasound probes in routine condition. During the first phase, the probe ( = 100) was sampled after a compl...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of clinical medicine 2019-10, Vol.8 (11), p.1776
Hauptverfasser: Pichon, Maxime, Lebail-Carval, Karine, Billaud, Geneviève, Lina, Bruno, Gaucherand, Pascal, Mekki, Yahia
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:This three-step study evaluated ultraviolet-C (UV-C) efficacy against human papillomavirus (HPV) found on vaginal ultrasound probes. The first two steps evaluated UV-C disinfection of vaginal ultrasound probes in routine condition. During the first phase, the probe ( = 100) was sampled after a complete cleaning and disinfection protocol, i.e., cleaning with chemically impregnated wipes, followed by UV-C. During the second phase, the probe ( = 47) was sampled after cleaning and UV-C. The final step consisted of applying mixes of HPV on a dedicated, covered probe ( = 15) then sampling the cover, the probe after removal of the cover, after cleaning, and after UV-C. HPV detection was performed using CLART HPV2 PCR (Genomica, Madrid, Spain). In the first phase, no probes were found to be positive for both DNA after UV-C. In the second phase, eight probes were found to be positive after cleaning (seven with human DNA and one with HPV) and negative after UV-C. In the final phase, one probe was found to be positive for HPV for each sample except after UV-C. Covers followed by a chemically impregnated wipe are not sufficient to ensure patient safety during vaginal ultrasound examinations. UV-C is effective in routine conditions against contaminations found on vaginal ultrasound probes, especially HPV.
ISSN:2077-0383
2077-0383
DOI:10.3390/jcm8111776