Radiosurgery for mesial temporal lobe epilepsy following ROSE trial guidelines — A planning comparison between Gamma Knife, Eclipse, and Brainlab
Purpose This study aims to compare stereotactic radiosurgery (SRS) planning of epilepsy that complies with Radiosurgery or Open Surgery for Epilepsy (ROSE) guidelines in GammaKnife, non‐coplanar conformal (NCC) plan in Eclipse, dynamic conformal arc (DCA) plan in Brainlab, and a volumetric modulated...
Gespeichert in:
Veröffentlicht in: | Journal of applied clinical medical physics 2019-10, Vol.20 (10), p.134-141 |
---|---|
Hauptverfasser: | , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Purpose
This study aims to compare stereotactic radiosurgery (SRS) planning of epilepsy that complies with Radiosurgery or Open Surgery for Epilepsy (ROSE) guidelines in GammaKnife, non‐coplanar conformal (NCC) plan in Eclipse, dynamic conformal arc (DCA) plan in Brainlab, and a volumetric modulated arc therapy (VMAT) plan in Eclipse.
Methods
Twenty plans targeting Mesial temporal lobe epilepsy (MTLE) was generated using GammaKnife, Eclipse with 20 NCC beams, Brainlab with 5 DCA, and Eclipse VMAT with 4 arcs observing ROSE trial guidelines. Multivariate analysis of variance and Wilcoxon signed‐rank test were used to compare dosimetric data of the plans and perform pairwise comparison, respectively.
Results
The plans obeyed the recommended prescription isodose volume (PIV) within 5.5–7.5 cc and maximum doses to brainstem, optic apparatus (OA) of 10 and 8 Gy, respectively, for a prescription dose of 24 Gy. The volumes of the target were in the range 4.0–7.4 cc. Mean PIV, maximum dose to brainstem, OA were 6.5 cc, 10 Gy, 7.9 Gy in GammaKnife; 7.2 cc, 6.1 Gy, 4.5 Gy in Eclipse NCC; 7.2 cc, 6.4 Gy, 5.7 Gy in Brainlab DCA; and 5.2 cc, 8.4 Gy, 6.1 Gy in Eclipse VMAT plans, respectively. Multivariate analysis of variance showed significant differences among the 4 SRS planning techniques (P‐values |
---|---|
ISSN: | 1526-9914 1526-9914 |