Evaluation of an Ultraviolet C (UVC) Light-Emitting Device for Disinfection of High Touch Surfaces in Hospital Critical Areas

Implementation of environmental cleaning and disinfection has been shown to reduce the incidences of healthcare-associated infections. The effect of an enhanced strategy for terminal room disinfection, applying the pulsed xenon-based ultraviolet light no-touch disinfection systems (PX-UVC) after the...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:International journal of environmental research and public health 2019-09, Vol.16 (19), p.3572
Hauptverfasser: Casini, Beatrice, Tuvo, Benedetta, Cristina, Maria Luisa, Spagnolo, Anna Maria, Totaro, Michele, Baggiani, Angelo, Privitera, Gaetano Pierpaolo
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Implementation of environmental cleaning and disinfection has been shown to reduce the incidences of healthcare-associated infections. The effect of an enhanced strategy for terminal room disinfection, applying the pulsed xenon-based ultraviolet light no-touch disinfection systems (PX-UVC) after the current standard operating protocol (SOP) was evaluated. In a teaching hospital, the effectiveness in reducing the total bacterial count (TBC) and in eliminating high-concern microorganisms was assessed on five high-touch surfaces in different critical areas, immediately pre- and post-cleaning and disinfection procedures (345 sampling sites). PX-UVC showed only 18% (15/85) of positive samples after treatment compared to 63% (72/115) after SOP. The effectiveness of PX-UVC was also observed in the absence of manual cleaning and application of a chemical disinfectant. According to the hygienic standards proposed by the Italian Workers Compensation Authority, 9 of 80 (11%) surfaces in operating rooms showed TBC ≥15 CFU/24 cm after the SOP, while all samples were compliant applying the SOP plus PX-UVC disinfection. (CD) spores and (KPC) were isolated only after the SOP. The implementation of the standard cleaning and disinfection procedure with the integration of the PX-UVC treatment had effective results in both the reduction of hygiene failures and in control environmental contamination by high-concern microorganisms.
ISSN:1660-4601
1661-7827
1660-4601
DOI:10.3390/ijerph16193572