Four‐year results of a minimally invasive transscleral glaucoma gel stent implantation in a prospective multi‐centre study
Importance The transscleral XEN Glaucoma Gel Microstent (XEN‐GGM, Allergan Plc., Parsippany, New Jersey) is implanted by a minimally invasive ab interno technique. Background The present study aims to assess the long‐term clinical outcomes in patients after XEN‐GGM implantation. Design This prospect...
Gespeichert in:
Veröffentlicht in: | Clinical & experimental ophthalmology 2019-07, Vol.47 (5), p.581-587 |
---|---|
Hauptverfasser: | , , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Importance
The transscleral XEN Glaucoma Gel Microstent (XEN‐GGM, Allergan Plc., Parsippany, New Jersey) is implanted by a minimally invasive ab interno technique.
Background
The present study aims to assess the long‐term clinical outcomes in patients after XEN‐GGM implantation.
Design
This prospective, non‐randomized, multi‐centred study was conducted in three countries (Austria, Canada and Germany).
Participants
Sixty‐four consecutive eyes of 64 patients with open angle glaucoma received the XEN‐GGM (63 μm) without Mitomycin C. Thirty‐five (55%) were solo procedures, and 29 (45%) were combined with cataract surgery.
Methods
Visits were planned at baseline, 6 months, 1, 2, 3 and 4 years postoperatively.
Main Outcome Measures
The main outcome measures were mean intraocular pressure (IOP), mean number of IOP lowering medication. Secondary outcome parameters were: visual acuity, visual fields and complete surgical failure (defined as presence of a secondary IOP lowering procedure or loss of light perception) at 4 years, postoperatively.
Results
Mean best‐medicated baseline IOP was 22.5 ± 4.2 mmHg and decreased significantly to 13.4 ± 3.1 mmHg 4 years postoperatively (−40%, n = 34, P < 0.001). Mean number of IOP lowering medication decreased significantly from 2.4 ± 1.3 preoperatively to 1.2 ± 1.3 (−50%, n = 34, P < 0.001) postoperatively. Visual field mean deviation showed no significant change between preoperative and postoperative examinations. Complete surgical failure rate per year was 10%.
Conclusions and Relevance
The XEN‐GGM resulted in lower IOP and a reduction in medications from baseline over 4 years of follow‐up. There was no detectable decrease in visual fields over the study. The surgical failure rate is comparable to other filtration surgeries. |
---|---|
ISSN: | 1442-6404 1442-9071 |
DOI: | 10.1111/ceo.13463 |