A Novel Toroidal-Flow Left Ventricular Assist Device Minimizes Blood Trauma: Implications of Improved Ventricular Assist Device Hemocompatibility

Continuous-flow left ventricular assist devices (LVADs) cause blood trauma that includes von Willebrand factor degradation, platelet activation, and subclinical hemolysis. Blood trauma contributes to bleeding, thrombosis, and stroke, which cause significant morbidity and mortality. The TORVAD (Windm...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:The Annals of thoracic surgery 2019-06, Vol.107 (6), p.1761-1767
Hauptverfasser: Bartoli, Carlo R., Hennessy-Strahs, Samson, Gohean, Jeff, Villeda, Maryann, Larson, Erik, Longoria, Raul, Kurusz, Mark, Acker, Michael A., Smalling, Richard
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Continuous-flow left ventricular assist devices (LVADs) cause blood trauma that includes von Willebrand factor degradation, platelet activation, and subclinical hemolysis. Blood trauma contributes to bleeding, thrombosis, and stroke, which cause significant morbidity and mortality. The TORVAD (Windmill Cardiovascular Systems, Inc, Austin, TX) is a first-of-its kind, toroidal-flow LVAD designed to minimize blood trauma. We tested the hypothesis that the TORVAD causes less blood trauma than the HeartMate II (Abbott Laboratories, Pleasanton, CA) LVAD. Whole human blood was circulated for 6 hours in ex vivo circulatory loops with a HeartMate II (n = 8; 10,000 rpm, 70 ± 6 mm Hg, 4.0 ± 0.1 L/min) or TORVAD (n = 6; 144 rpm, 72 ± 0.0 mm Hg, 4.3 ± 0.0 L/min). von Willebrand factor degradation was quantified with electrophoresis and immunoblotting. Platelet activation was quantified by cluster of differentiation (CD) 41/61 enzyme-linked immunosorbent assay (ELISA). Hemolysis was quantified by plasma free hemoglobin ELISA. The TORVAD caused significantly less degradation of high-molecular-weight von Willebrand factor multimers (−10% ± 1% vs −21% ± 1%, p 
ISSN:0003-4975
1552-6259
DOI:10.1016/j.athoracsur.2018.11.053