Tolerance levels of mass density for CT number calibration in photon radiation therapy

Computed tomography (CT) data are required to calculate the dose distribution in a patient’s body. Generally, there are two CT number calibration methods for commercial radiotherapy treatment planning system (RTPS), namely CT number‐relative electron density calibration (CT‐RED calibration) and CT n...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of applied clinical medical physics 2019-06, Vol.20 (6), p.45-52
Hauptverfasser: Nakao, Minoru, Ozawa, Shuichi, Yogo, Katsunori, Miura, Hideharu, Yamada, Kiyoshi, Hosono, Fumika, Hayata, Masahiro, Miki, Kentaro, Nakashima, Takeo, Ochi, Yusuke, Kawahara, Daisuke, Morimoto, Yoshiharu, Yoshizaki, Toru, Nozaki, Hiroshige, Habara, Kosaku, Nagata, Yasushi
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Computed tomography (CT) data are required to calculate the dose distribution in a patient’s body. Generally, there are two CT number calibration methods for commercial radiotherapy treatment planning system (RTPS), namely CT number‐relative electron density calibration (CT‐RED calibration) and CT number‐mass density calibration (CT‐MD calibration). In a previous study, the tolerance levels of CT‐RED calibration were established for each tissue type. The tolerance levels were established when the relative dose error to local dose reached 2%. However, the tolerance levels of CT‐MD calibration are not established yet. We established the tolerance levels of CT‐MD calibration based on the tolerance levels of CT‐RED calibration. In order to convert mass density (MD) to relative electron density (RED), the conversion factors were determined with adult reference computational phantom data available in the International Commission on Radiological Protection publication 110 (ICRP‐110). In order to validate the practicability of the conversion factor, the relative dose error and the dose linearity were validated with multiple RTPSes and dose calculation algorithms for two groups, namely, CT‐RED calibration and CT‐MD calibration. The tolerance levels of CT‐MD calibration were determined from the tolerance levels of CT‐RED calibration with conversion factors. The converted RED from MD was compared with actual RED calculated from ICRP‐110. The conversion error was within ±0.01 for most standard organs. It was assumed that the conversion error was sufficiently small. The relative dose error difference for two groups was less than 0.3% for each tissue type. Therefore, the tolerance levels for CT‐MD calibration were determined from the tolerance levels of CT‐RED calibration with the conversion factors. The MD tolerance levels for lung, adipose/muscle, and cartilage/spongy‐bone corresponded to ±0.044, ±0.022, and ±0.045 g/cm3, respectively. The tolerance levels were useful in terms of approving the CT‐MD calibration table for clinical use.
ISSN:1526-9914
1526-9914
DOI:10.1002/acm2.12601