Full density fluorescence molecular tomography (FD-FMT) based on a dichroic mirror

We present a novel method called full density fluorescence molecular tomography (FD-FMT) that can considerably improve the performance of conventional FMT. By converting each source (or detector) to a detector (or source) through the use of a dichroic mirror, FD-FMT not only increases the amount of...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Applied optics 2018-09, Vol.57 (27), p.7938-7941
Hauptverfasser: Yang, Hao, Dai, Xianjin, Jiang, Huabei
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We present a novel method called full density fluorescence molecular tomography (FD-FMT) that can considerably improve the performance of conventional FMT. By converting each source (or detector) to a detector (or source) through the use of a dichroic mirror, FD-FMT not only increases the amount of optical projections by more than fourfold (compared to conventional FMT) to achieve high-resolution image reconstruction, but also offers the possibility to realize miniaturized FMT systems.