Managing Arrow-Trerotola Percutaneous Thrombolytic Device Complications
Abstract Dialysis access interventions are frequently performed by interventional radiologists. Several commercially available percutaneous thrombolytic devices can help restore patency to thrombosed arteriovenous access circuits. The Arrow-Trerotola Percutaneous Thrombolytic Device is one such devi...
Gespeichert in:
Veröffentlicht in: | Seminars in interventional radiology 2019-06, Vol.36 (2), p.091-096 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Abstract
Dialysis access interventions are frequently performed by interventional radiologists. Several commercially available percutaneous thrombolytic devices can help restore patency to thrombosed arteriovenous access circuits. The Arrow-Trerotola Percutaneous Thrombolytic Device is one such device, and has a long track record of safe and effective use. However, like any medical device, complications can occur during its use. This article describes three complications and associated management strategies utilizing fundamental interventional radiology techniques of balloon tamponade, stent placement, and snare mediated foreign body retrieval. |
---|---|
ISSN: | 0739-9529 1098-8963 |
DOI: | 10.1055/s-0039-1688421 |