Surface cytometer for fluorescent detection and growth monitoring of bacteria over a large field-of-view
Monitoring the early onset of bacterial film formation is critical in many clinical, environmental, and food quality control applications. We built a small inexpensive optical surface cytometer, in contrast with bulk spectroscopic methods, around a light-emitting diode (LED) and a complementary meta...
Gespeichert in:
Veröffentlicht in: | Biomedical optics express 2019-04, Vol.10 (4), p.2101-2116 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Monitoring the early onset of bacterial film formation is critical in many clinical, environmental, and food quality control applications. We built a small inexpensive optical surface cytometer, in contrast with bulk spectroscopic methods, around a light-emitting diode (LED) and a complementary metal-oxide-semiconductor (CMOS) image sensor. It is designed to offer a large field-of-view of 200 mm
and a large depth-of-field of 2-3 mm to overcome the limitations of routine methods like spectrophotometry and fluorescence microscopy. It provides a direct measurement without the need for complex image post-processing with a limit-of-detection around 10
cells/mm
, which is competitive with other similar yet more complex devices already available. |
---|---|
ISSN: | 2156-7085 2156-7085 |
DOI: | 10.1364/BOE.10.002101 |