High-precision analysis of 182W/184W and 183W/184W by negative thermal ionization mass spectrometry: Per-integration oxide corrections using measured 18O/16O
[Display omitted] •Oxygen isotopic compositions are measured in situ for every integration, allowing correction for in-run oxygen fractionation.•Average of newly determined oxygen isotopic compositions is used to make first-order oxide corrections to interfering oxides.•Long-term external precisions...
Gespeichert in:
Veröffentlicht in: | International journal of mass spectrometry 2017-03, Vol.414, p.80-86 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | [Display omitted]
•Oxygen isotopic compositions are measured in situ for every integration, allowing correction for in-run oxygen fractionation.•Average of newly determined oxygen isotopic compositions is used to make first-order oxide corrections to interfering oxides.•Long-term external precisions are 5.7ppm and 3.7ppm (2SD) for 182W/184W when mass bias corrected using 186W/184W and 186W/183W, respectively.•Long-term external precision is 6.6ppm (2SD) for 183W/184W when mass bias corrected using 186W/184W.•Measurement of IVB iron meteorite Skookum using this method is shown to be accurate and precise.
Here we describe a new analytical technique for the high-precision measurement of 182W/184W and 183W/184W using negative thermal ionization mass spectrometry (N-TIMS). We improve on the recently reported method of Trinquier et al. (2016), which described using Faraday cup collectors coupled with amplifiers utilizing 1013Ω resistors to continuously monitor the 18O/16O of WO3− and make per-integration oxide corrections. In our study, we report and utilize a newly measured oxygen mass fractionation line, as well as average 17O/16O and 18O/16O, which allow for more accurate per-integration oxide interference corrections. We also report a Faraday cup and amplifier configuration that allows 18O/16O to be continuously monitored for WO3− and ReO3−, both of which are ionized during analyses of W using Re ribbon. The long-term external precision of 182W/184W is 5.7ppm and 3.7ppm (2SD) when mass bias corrected using 186W/184W and 186W/183W, respectively. For 183W/184W mass bias is corrected using 186W/184W, yielding a long-term external precision of 6.6ppm. An observed, correlated variation in 182W/184W and 183W/184W, when mass bias corrected using 186W/184W, is most likely the result of Faraday cup degradation over months-long intervals. |
---|---|