Hollow Fiber Oxygenator Composition Has a Significant Impact on Failure Rates in Neonates on Extracorporeal Membrane Oxygenation: A Retrospective Analysis

Abstract In extracorporeal life support (ECLS), there are two main types of oxygenators in clinical use for neonates: polymethylpentene (PMP) hollow fiber and polypropylene (PP) hollow fiber. A retrospective study was performed on neonates ( n  = 44) who had undergone ECLS for noncardiac indications...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of pediatric intensive care 2018-03, Vol.7 (1), p.007-013
Hauptverfasser: Daniel, John M., Bernard, Philip A., Skinner, Sean C., Bhandary, Prasad, Ruzic, Ana, Bacon, Matthew K., Ballard, Hubert O.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Abstract In extracorporeal life support (ECLS), there are two main types of oxygenators in clinical use for neonates: polymethylpentene (PMP) hollow fiber and polypropylene (PP) hollow fiber. A retrospective study was performed on neonates ( n  = 44) who had undergone ECLS for noncardiac indications from 2009 to 2015. Between the two groups (PMP n  = 21, PP n  = 23), the PP oxygenators failed 91% of the time, whereas the PMP oxygenators failed 43% of the time ( p  
ISSN:2146-4618
2146-4626
DOI:10.1055/s-0037-1599150