Clinical Interpretation of Peripheral Pulse Oximeters Labeled “Not for Medical Use”

Abstract The purpose of our study was to clarify limitations of off-label use for low cost nonmedical use (NMU) pulse oximeters by primary care providers. These devices are widely marketed over the Internet and in drugstores but are not intended for medical use or reviewed by the Food and Drug Admin...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Annals of family medicine 2018-11, Vol.16 (6), p.552-554
Hauptverfasser: Hudson, Arlene J., MD, MA, Benjamin, John, MD, Jardeleza, Timothy, MD, Bergstrom, Curt, MD, Cronin, William, MD, Mendoza, Mario, MD, MS, Schultheis, Lex, MD, PhD
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Abstract The purpose of our study was to clarify limitations of off-label use for low cost nonmedical use (NMU) pulse oximeters by primary care providers. These devices are widely marketed over the Internet and in drugstores but are not intended for medical use or reviewed by the Food and Drug Administration (FDA). Our study compared oxygen saturation (SpO2 ) in patients from 1 medical use (MU) pulse oximeter to 8 NMU pulse oximeters. Measured arterial oxygenation (SaO2 ) was compared with SpO2 when available. In patients who were normoxic (SpO2 ≥90%), all oximeters exhibited similar readings. This finding suggests that NMU pulse oximeters may be able to rule out hypoxemia in clinical settings.
ISSN:1544-1709
1544-1717
DOI:10.1370/afm.2317