Effectiveness of a New Single‐Needle Single‐Pump Dialysis System with Simultaneous Monitoring of Dialysis Dose
To date, single‐needle (SN) hemodialysis (HD) requires a dialysis machine equipped with two blood pumps—one controlling arterial blood flow (Qb) and one controlling venous Qb. B. Braun has developed an innovative single‐pump SN HD system. Therefore, usability is improved by reducing complexity. The...
Gespeichert in:
Veröffentlicht in: | Artificial organs 2018-08, Vol.42 (8), p.814-823 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | To date, single‐needle (SN) hemodialysis (HD) requires a dialysis machine equipped with two blood pumps—one controlling arterial blood flow (Qb) and one controlling venous Qb. B. Braun has developed an innovative single‐pump SN HD system. Therefore, usability is improved by reducing complexity. The aim of this study was to compare dialysis parameters of the new single‐pump SN HD system with a double‐pump SN HD system available on the market (Fresenius Medical Care [FMC] 5008). In this two‐armed crossover study, patients were randomized into two groups (B. Braun ‐ FMC/FMC ‐ B. Braun). Study period was 2 weeks (6 HD sessions) for each SN HD system. Both B. Braun and FMC dialysis machines were operated in the single‐needle auto mode. With the FMC dialysis machines, Qb was optimized manually, whereas for B. Braun machines it was optimized automatically using the auto‐mode functionality. A phase volume of 25 mL, treatment time, needle type and size, and dialyzer type and size were kept constant per patient throughout the study. Due to technical prerequisites in the SN mode, online dialysis adequacy (Kt/V: K ‐ dialyzer clearance of urea; t ‐ dialysis time; V ‐ volume of distribution of urea) monitoring could only be performed in the B. Braun group. Twelve HD patients (5 male/7 female, mean age 75.5 ± 8.8 years, mean time on dialysis 4.97 ± 3.86 years, 3× weekly HD) were enrolled. Total number of treatments performed: n = 132 (65 B. Braun, 67 FMC) and the mean online Kt/V value in the B. Braun group was 1.26 ± 0.29 (n = 63). Mean dialysis time per session: B. Braun 253.4 ± 19.9 min, FMC 251.6 ± 18.8 min. Mean phase volume: B. Braun 25.1 ± 0.2 mL, FMC 25.4 ± 3.1 mL. Mean cumulated blood volume (CBV): B. Braun 55.0 ± 5.5 L, FMC 40.5 ± 5.9 L (P |
---|---|
ISSN: | 0160-564X 1525-1594 |
DOI: | 10.1111/aor.13149 |