Analysis of the long-term stability of the output of electron beams generated by the Novac11™ IORT accelerator

The aim of the study was to analyze the long-term stability of electron beams generated by the Novac11™ IORT accelerator. Novac11™ (NRT®) is a mobile electron accelerator designed to irradiate small areas of tissue, up to 10 cm in diameter, with electron beams during surgical procedures. It is chara...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Reports of practical oncology and radiotherapy 2018-09, Vol.23 (5), p.341-345
Hauptverfasser: Bijok, Michał, Gruszczynska, Ewelina, Kowalczyk, Adam, Sikorska, Katarzyna, Walewska, Agnieszka, Kukolowicz, Pawel F.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The aim of the study was to analyze the long-term stability of electron beams generated by the Novac11™ IORT accelerator. Novac11™ (NRT®) is a mobile electron accelerator designed to irradiate small areas of tissue, up to 10 cm in diameter, with electron beams during surgical procedures. It is characterized by a great mobility guaranteed by a number of degrees of freedom enabling irradiation in the conditions of an operating theatre. Over the period of January 2013 and September 2016, the measurement sessions of the output of clinically used beam qualities (6, 8 and 10 MeV) were carried out 41 times. Because of the unsatisfactory long-term stability, an extra procedure of tuning of the magnetron, suggested by the manufacturer, was introduced in October 2015, 15 measurements were performed since then. The output of the Novac11™ accelerator was measured in the reference conditions recommended by the IAEA Report 398, the measurements of the charge in the ionization chamber at the reference depth were carried out with a Dose1™ electrometer and a plane-parallel chamber PPC05™ from IBA®. The introduction of the tuning of the magnetron procedure resulted in satisfactory long-term stability of the measured outputs below 2%. After the introduction of the STV parameter tuning procedure, the long-term stability of the Novac11™ output increased considerably and is within the values declared by the manufacturer.
ISSN:1507-1367
2083-4640
DOI:10.1016/j.rpor.2018.07.003