Effects of bed compression on protein separation on gel filtration chromatography at bench and pilot scale
BACKGROUND Poorly packed chromatography columns are known to reduce drastically the column efficiency and produce broader peaks. Controlled bed compression has been suggested to be a useful approach for solving this problem. Here the relationship between column efficiency and resolution of protein s...
Gespeichert in:
Veröffentlicht in: | Journal of chemical technology and biotechnology (1986) 2018-07, Vol.93 (7), p.1959-1965 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | BACKGROUND
Poorly packed chromatography columns are known to reduce drastically the column efficiency and produce broader peaks. Controlled bed compression has been suggested to be a useful approach for solving this problem. Here the relationship between column efficiency and resolution of protein separation are examined when preparative chromatography media were compressed using mechanical and hydrodynamic methods. Sepharose CL‐6B, an agarose based size exclusion media was examined at bench and pilot scale. The asymmetry and height equivalent of a theoretical plate (HETP) was determined by using 2% v/v acetone, whereas the void volume and intraparticle porosity (ϵp) were estimated by using blue dextran. A protein mixture of ovalbumin (chicken), bovine serum albumin (BSA) and γ'‐ globulin (bovine) with molecular weights of 44, 67 and 158 kDa, respectively, were used as a ‘model’ separation challenge.
RESULTS
Mechanical compression achieved a reduction in plate height for the column with a concomitant improvement in asymmetry. Furthermore, the theoretical plate height decreased significantly with mechanical compression resulting in a 40% improvement in purity compared with uncompressed columns at the most extreme conditions of compression used.
CONCLUSION
The results suggest that the mechanical bed compression of Sepharose CL‐6B can be used to improve the resolution of protein separation. © 2017 The Authors. Journal of Chemical Technology & Biotechnology published by John Wiley & Sons Ltd on behalf of Society of Chemical Industry. |
---|---|
ISSN: | 0268-2575 1097-4660 |
DOI: | 10.1002/jctb.5411 |