Evaluation of Performance Parameters of the Disposable Flexible Ureterorenoscope (LITHOVUE) in Patients with Renal Stones: A Prospective, Observational, Single-arm, Multicenter Study

We investigated performance parameters of disposable flexible ureterorenoscopy (LITHOVUE) in patients with renal stones in a prospective, multicenter, observational study. Sixty two patients who underwent ureterorenoscopic surgery by LITHOVUE were included. Surgeons using a numerical scale, evaluate...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Scientific reports 2018-06, Vol.8 (1), p.9795-6, Article 9795
Hauptverfasser: Cho, Sung Yong, Lee, Joo Yong, Shin, Dong Gil, Seo, Ill Young, Yoo, Sangjun, Park, Hyung Keun
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We investigated performance parameters of disposable flexible ureterorenoscopy (LITHOVUE) in patients with renal stones in a prospective, multicenter, observational study. Sixty two patients who underwent ureterorenoscopic surgery by LITHOVUE were included. Surgeons using a numerical scale, evaluated the maneuverability of the scopes and general perceptions of visibility related to the monitor and irrigation systems. General pain and fatigue scores were evaluated and compared to existing scopes. Mean patient age was 57.3 years and stone size was 15.4 mm. Clinical success of overall stone removal was 82.3%. Complications occurred in 4 cases of the Clavien classification grade I in a single case and II in 3 cases. Comparative parameters of maneuverability, perception of the monitor system and perception of the irrigation channel were 2.5, 2.5, and 3.0, respectively. The most favorable evaluation of physical strain was about ‘shoulder fatigue’ and ‘hand fatigue’. Unfavorable evaluations were recorded for ‘wrist stiffness’ and ‘thumb fatigue’. Maximal deflection angles of 270° were preserved in 53 cases (85.5%). No pre-stenting procedure and the longer operative time were significant predictors of poorly-preserved deflection angles
ISSN:2045-2322