Patient movement characteristics and the impact on CBCT image quality and interpretability

To assess the impact of patient movement characteristics and metal/radiopaque materials in the field-of-view (FOV) on CBCT image quality and interpretability. 162 CBCT examinations were performed in 134 consecutive (i.e. prospective data collection) patients (age average: 27.2 years; range: 9-73). A...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Dento-maxillo-facial radiology 2018-01, Vol.47 (1), p.20170216-20170216
Hauptverfasser: Spin-Neto, Rubens, Costa, Cláudio, Salgado, Daniela Mra, Zambrana, Nataly Rm, Gotfredsen, Erik, Wenzel, Ann
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:To assess the impact of patient movement characteristics and metal/radiopaque materials in the field-of-view (FOV) on CBCT image quality and interpretability. 162 CBCT examinations were performed in 134 consecutive (i.e. prospective data collection) patients (age average: 27.2 years; range: 9-73). An accelerometer-gyroscope system registered patient's head position during examination. The threshold for movement definition was set at ≥0.5-mm movement distance based on accelerometer-gyroscope recording. Movement complexity was defined as uniplanar/multiplanar. Three observers scored independently: presence of stripe (i.e. streak) artefacts (absent/"enamel stripes"/"metal stripes"/"movement stripes"), overall unsharpness (absent/present) and image interpretability (interpretable/not interpretable). Kappa statistics assessed interobserver agreement. χ tests analysed whether movement distance, movement complexity and metal/radiopaque material in the FOV affected image quality and image interpretability. Relevant risk factors (p ≤ 0.20) were entered into a multivariate logistic regression analysis with "not interpretable" as the outcome. Interobserver agreement for image interpretability was good (average = 0.65). Movement distance and presence of metal/radiopaque materials significantly affected image quality and interpretability. There were 22-28 cases, in which the observers stated the image was not interpretable. Small movements (i.e.
ISSN:0250-832X
1476-542X
DOI:10.1259/dmfr.20170216