Are there increased periprocedural complications with the MRI-conditional Medtronic Revo SureScan Pacing System?: A meta-analysis
Background The use of magnetic resonance imaging (MRI)-conditional permanent pacemakers has increased significantly. In this meta-analysis, we examine the safety of MRI-conditional pacing systems in comparison with conventional systems. Methods An electronic search was performed using major database...
Gespeichert in:
Veröffentlicht in: | Netherlands heart journal 2018-05, Vol.26 (5), p.233-239 |
---|---|
Hauptverfasser: | , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Background
The use of magnetic resonance imaging (MRI)-conditional permanent pacemakers has increased significantly. In this meta-analysis, we examine the safety of MRI-conditional pacing systems in comparison with conventional systems.
Methods
An electronic search was performed using major databases, including studies that compared the outcomes of interest between patients receiving MRI-conditional pacemakers (MRI group) versus conventional pacemakers (control group).
Results
Six studies (5 retrospective and 1 prospective non-randomised) involving 2,118 adult patients were identified. The MRI-conditional pacemakers, deployed in 969 patients, were all from a single manufacturer (Medtronic Pacing System with 5086 leads). The rate of pacemaker lead dislodgement (atrial and ventricular) was significantly higher in the MRI group (3% vs. 1%, OR 2.47 (95% CI 1.26; 4.83),
p
= 0.008). The MRI group had a significantly higher rate of pericardial complications (2% vs. 1%, OR 4.23 (95% CI 1.18; 15.10),
p
= 0.03) and a numerically higher overall complication rate in comparison with the conventional group (6% vs. 3%, OR 2.02 (95% CI 0.88; 4.66),
p
= 0.10) but this was not statistically significant.
Conclusions
In this meta-analysis, the rates of pacemaker lead dislodgement and pericardial complications were significantly higher with the Medtronic MRI-conditional pacing system. |
---|---|
ISSN: | 1568-5888 1876-6250 |
DOI: | 10.1007/s12471-018-1086-4 |