Patient radiation exposure during general fluoroscopy examinations

The purpose of this study was to assess the level of patient radiation dose received in general fluoroscopy examinations, compare the findings with the international diagnostic reference levels (IDRLs), and establish the initial institutional (local) LDRLs. A comprehensive survey was conducted for g...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of applied clinical medical physics 2014-03, Vol.15 (2), p.262-270
Hauptverfasser: Wambani, Jeska S., Korir, Geoffrey K., Tries, Mark A., Korir, Ian K., Sakwa, Jedidah M.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The purpose of this study was to assess the level of patient radiation dose received in general fluoroscopy examinations, compare the findings with the international diagnostic reference levels (IDRLs), and establish the initial institutional (local) LDRLs. A comprehensive survey was conducted for general fluoroscopy examinations using the medical records of a Radiology Department of a leading regional hospital over a period close to one year. The cumulative reference point air kerma (Ka,r), kerma area product (KAP) and fluoroscopy time (FT) were recorded for six hundred and fifty (30% pediatric and 70% adult) patients undergoing routine fluoroscopy examinations using X‐ray equipment with built‐in integrated dose measuring system. Results which were obtained for adult general fluoroscopy indicated that 83% and 33% were below the IDRLs for KAP and fluoroscopy time, respectively. In children, 60% were found to be below the only available KAP diagnostic reference levels. Local diagnostic reference levels (LDRLs) have been proposed with respect to the missing DRLs for the Ka r, KAP, and fluoroscopy time. The majority of the examinations in the study were performed with longer fluoroscopy time, patient dose values per examination type were found to be broad and the mean values above the international diagnostic reference levels. This calls for proper and improved training and radiation protection skills for the responsible personnel, especially the equipment operators. PACS numbers: 87.53.Bn, 87.59.C‐, 87.59.cf, 87.53.Bn, 87.50.‐a, 87.53.‐j
ISSN:1526-9914
1526-9914
DOI:10.1120/jacmp.v15i2.4555