Evaluation of a newly developed quantitative determination kit for tumor marker CA15‐3 with chemiluminescent assay

Background Tumor marker carbohydrate antigen 15‐3 (CA15‐3) is used as a biomarker to aid to diagnose and monitor the prognosis of breast cancer patients. A new quantitative determination kit for CA15‐3 with chemiluminescent assay was developed by Xiamen InnoDx Biotech Co., Ltd, China. Therefore, we...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of clinical laboratory analysis 2018-01, Vol.32 (1), p.n/a
Hauptverfasser: Li, Peihua, Ye, Huiming, Liu, Jiangwu, Jin, Hongwei, Lin, Yongzhi, Yan, Shuidi, Yu, Yang, Gao, Lei, Xu, Feihai, Zhang, Zhongying
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Background Tumor marker carbohydrate antigen 15‐3 (CA15‐3) is used as a biomarker to aid to diagnose and monitor the prognosis of breast cancer patients. A new quantitative determination kit for CA15‐3 with chemiluminescent assay was developed by Xiamen InnoDx Biotech Co., Ltd, China. Therefore, we conducted the report to evaluate the performance of the kit. Methods According to the “Guiding principles on performance analysis of diagnostic reagents in vitro”, the calibration curve, limit of detection, reportable range, accuracy, precision, anti‐interference capability, cross‐reaction and comparison by measuring EDTA plasma and serum were carried out. In addition, the kit was performed in parallel to electrochemiluminescence immunoassay kit (Roche) to analyze the correlation between the two kits. Results Regression equation of calibration curve of the kit was Y=0.7914X+4.1032 (R2=.990). Limit of detection was 0.0347 U/mL. The reportable range was 0.5‐2400 U/mL. Recovery ratio was 100.0%‐104.8%. Coefficient of variations (CVs) of within‐run and between‐run were 4.8%‐7.6% and 5.8%‐7.4% respectively. No remarkable interferences (all Bias% were less than ±10%) were detected when samples contained hemoglobin ≤183.8 μmol/L, bilirubin ≤340 μmol/L, triglyceride ≤18.1 mmol/L, or rheumatoid factor ≤400 U/mL. No cross‐reaction was present in the kit. Moreover, compared with the results from electrochemiluminescence immunoassay kit (Roche) in 345 serum samples, there was a satisfied correlation coefficient of 0.977 (P
ISSN:0887-8013
1098-2825
DOI:10.1002/jcla.22158