Rapid Detection of Carbapenemase Production in Enterobacteriaceae by Use of a Modified Paper Strip Carba NP Method

Rapid and accurate detection of carbapenemase-producing (CPE) is important for preventing their spread in health care settings. We compared the performance of the Carba NP (CNP) test using the CLSI tube method with that using a modified paper strip method for the detection of carbapenemases in 390 i...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of clinical microbiology 2018-01, Vol.56 (1)
Hauptverfasser: Ho, Pak-Leung, Wang, Ya, Wing-Sze Tse, Cindy, Fung, Kitty Sau-Chun, Cheng, Vincent Chi-Chung, Lee, Rodney, To, Wing-Kin, Lai, Raymond Wai-Man, Luk, Wei-Kwang, Que, Tak-Lun, Tsang, Dominic Ngai-Chong
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Rapid and accurate detection of carbapenemase-producing (CPE) is important for preventing their spread in health care settings. We compared the performance of the Carba NP (CNP) test using the CLSI tube method with that using a modified paper strip method for the detection of carbapenemases in 390 isolates. The isolates were identified by Hong Kong's carbapenem-resistant surveillance program in 2016 and comprised 213 CPE and 177 carbapenemase-negative isolates. Molecular genotype was used as the reference. The test results were read at different time points for the CLSI method (1 min, 5 min, 1 h, and 2 h) and strip method (1 min and 5 min). The strip CNP and CLSI CNP tests correctly detect carbapenemase production in 93% and 93% of KPC producers, 100% and 38% of IMI producers, 94% and 85% of IMP producers, 98% and 90% of NDM producers, and 29% and 12% of OXA producers, respectively. Overall, the strip method has superior sensitivity to the CLSI method (86% versus 75%, respectively; < 0.001, McNemar test). The specificity of both methods was 100%. By the CLSI method, 27%, 14%, 29%, and 6% of the CPE isolates were positive at 1 min, 5 min, 1 h, and 2 h, respectively. In contrast, by the strip method, 76% of the CPE isolates were positive at 1 min, and an additional 10% were positive at 5 min. In conclusion, the Carba NP test by use of the modified strip method has a higher sensitivity and a shorter assay time than that those by use of the CLSI tube method.
ISSN:0095-1137
1098-660X
DOI:10.1128/JCM.01110-17