Calibration and error analysis of metal‐oxide‐semiconductor field‐effect transistor dosimeters for computed tomography radiation dosimetry

Purpose Metal‐oxide‐semiconductor field‐effect transistors (MOSFETs) serve as a helpful tool for organ radiation dosimetry and their use has grown in computed tomography (CT). While different approaches have been used for MOSFET calibration, those using the commonly available 100 mm pencil ionizatio...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Medical physics (Lancaster) 2017-12, Vol.44 (12), p.6589-6602
Hauptverfasser: Trattner, Sigal, Prinsen, Peter, Wiegert, Jens, Gerland, Elazar‐Lars, Shefer, Efrat, Morton, Tom, Thompson, Carla M., Yagil, Yoad, Cheng, Bin, Jambawalikar, Sachin, Al‐Senan, Rani, Amurao, Maxwell, Halliburton, Sandra S., Einstein, Andrew J.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Purpose Metal‐oxide‐semiconductor field‐effect transistors (MOSFETs) serve as a helpful tool for organ radiation dosimetry and their use has grown in computed tomography (CT). While different approaches have been used for MOSFET calibration, those using the commonly available 100 mm pencil ionization chamber have not incorporated measurements performed throughout its length, and moreover, no previous work has rigorously evaluated the multiple sources of error involved in MOSFET calibration. In this paper, we propose a new MOSFET calibration approach to translate MOSFET voltage measurements into absorbed dose from CT, based on serial measurements performed throughout the length of a 100‐mm ionization chamber, and perform an analysis of the errors of MOSFET voltage measurements and four sources of error in calibration. Methods MOSFET calibration was performed at two sites, to determine single calibration factors for tube potentials of 80, 100, and 120 kVp, using a 100‐mm‐long pencil ion chamber and a cylindrical computed tomography dose index (CTDI) phantom of 32 cm diameter. The dose profile along the 100‐mm ion chamber axis was sampled in 5 mm intervals by nine MOSFETs in the nine holes of the CTDI phantom. Variance of the absorbed dose was modeled as a sum of the MOSFET voltage measurement variance and the calibration factor variance, the latter being comprised of three main subcomponents: ionization chamber reading variance, MOSFET‐to‐MOSFET variation and a contribution related to the fact that the average calibration factor of a few MOSFETs was used as an estimate for the average value of all MOSFETs. MOSFET voltage measurement error was estimated based on sets of repeated measurements. The calibration factor overall voltage measurement error was calculated from the above analysis. Results Calibration factors determined were close to those reported in the literature and by the manufacturer (~3 mV/mGy), ranging from 2.87 to 3.13 mV/mGy. The error σV of a MOSFET voltage measurement was shown to be proportional to the square root of the voltage V: σV=cV where c = 0.11 mV. A main contributor to the error in the calibration factor was the ionization chamber reading error with 5% error. The usage of a single calibration factor for all MOSFETs introduced an additional error of about 5–7%, depending on the number of MOSFETs that were used to determine the single calibration factor. The expected overall error in a high‐dose region (~30 mGy) was estimated to be a
ISSN:0094-2405
2473-4209
DOI:10.1002/mp.12592