Use of a novel two‐dimensional ionization chamber array for pencil beam scanning proton therapy beam quality assurance

The need to accurately and efficiently verify both output and dose profiles creates significant challenges in quality assurance of pencil beam scanning (PBS) proton delivery. A system for PBS QA has been developed that combines a new two‐dimensional ionization chamber array in a waterproof housing t...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of applied clinical medical physics 2015-05, Vol.16 (3), p.270-276
Hauptverfasser: Lin, Liyong, Kang, Minglei, Solberg, Timothy D., Mertens, Thierry, Baumer, Christian, Ainsley, Christopher G., McDonough, James E.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The need to accurately and efficiently verify both output and dose profiles creates significant challenges in quality assurance of pencil beam scanning (PBS) proton delivery. A system for PBS QA has been developed that combines a new two‐dimensional ionization chamber array in a waterproof housing that is scanned in a water phantom. The MatriXX PT has the same detector array arrangement as the standard MatriXXEvolution but utilizes a smaller 2 mm plate spacing instead of 5 mm. Because the bias voltage of the MatriXX PT and Evolution cannot be changed, PPC40 and FC65‐G ionization chambers were used to assess recombination effects. The PPC40 is a parallel plate chamber with an electrode spacing of 2 mm, while the FC65‐G is a Farmer chamber FC65‐G with an electrode spacing of 2.8 mm. Three bias voltages (500, 200, and 100 V) were used for both detectors to determine which radiation type (continuous, pulse or pulse‐scanned beam) could closely estimate Pion from the ratios of charges collected. In comparison with the MatriXXEvolution, a significant improvement in measurement of absolute dose with the MatriXX PT was observed. While dose uncertainty of the MatriXXEvolution can be up to 4%, it is
ISSN:1526-9914
1526-9914
DOI:10.1120/jacmp.v16i3.5323