Evaluation of a combined respiratory‐gating system comprising the TrueBeam linear accelerator and a new real‐time tumor‐tracking radiotherapy system: a preliminary study
A combined system comprising the TrueBeam linear accelerator and a new real‐time, tumor‐tracking radiotherapy system, SyncTraX, was installed in our institution. The goals of this study were to assess the capability of SyncTraX in measuring the position of a fiducial marker using color fluoroscopic...
Gespeichert in:
Veröffentlicht in: | Journal of applied clinical medical physics 2016-07, Vol.17 (4), p.202-213 |
---|---|
Hauptverfasser: | , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A combined system comprising the TrueBeam linear accelerator and a new real‐time, tumor‐tracking radiotherapy system, SyncTraX, was installed in our institution. The goals of this study were to assess the capability of SyncTraX in measuring the position of a fiducial marker using color fluoroscopic images, and to evaluate the dosimetric and geometric accuracy of respiratory‐gated radiotherapy using this combined system for the simple geometry. For the fundamental evaluation of respiratory‐gated radiotherapy using SyncTraX, the following were performed: 1) determination of dosimetric and positional characteristics of sinusoidal patterns using a motor‐driven base for several gating windows; 2) measurement of time delay using an oscilloscope; 3) positional verification of sinusoidal patterns and the pattern in the case of a lung cancer patient; 4) measurement of the half‐value layer (HVL in mm AL), effective kVp, and air kerma, using a solid‐state detector for each fluoroscopic condition, to determine the patient dose. The dose profile in a moving phantom with gated radiotherapy having a gating window ≤4 mm was in good agreement with that under static conditions for each photon beam. The total time delay between TrueBeam and SyncTraX was |
---|---|
ISSN: | 1526-9914 1526-9914 |
DOI: | 10.1120/jacmp.v17i4.6114 |