ConoSurf: Open‐source 3D scanning system based on a conoscopic holography device for acquiring surgical surfaces
Background A difficulty in computer‐assisted interventions is acquiring the patient's anatomy intraoperatively. Standard modalities have several limitations: low image quality (ultrasound), radiation exposure (computed tomography) or high costs (magnetic resonance imaging). An alternative appro...
Gespeichert in:
Veröffentlicht in: | The international journal of medical robotics + computer assisted surgery 2017-09, Vol.13 (3), p.n/a |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Background
A difficulty in computer‐assisted interventions is acquiring the patient's anatomy intraoperatively. Standard modalities have several limitations: low image quality (ultrasound), radiation exposure (computed tomography) or high costs (magnetic resonance imaging). An alternative approach uses a tracked pointer; however, the pointer causes tissue deformation and requires sterilizing. Recent proposals, utilizing a tracked conoscopic holography device, have shown promising results without the previously mentioned drawbacks.
Methods
We have developed an open‐source software system that enables real‐time surface scanning using a conoscopic holography device and a wide variety of tracking systems, integrated into pre‐existing and well‐supported software solutions.
Results
The mean target registration error of point measurements was 1.46 mm. For a quick guidance scan, surface reconstruction improved the surface registration error compared with point‐set registration.
Conclusions
We have presented a system enabling real‐time surface scanning using a tracked conoscopic holography device. Results show that it can be useful for acquiring the patient's anatomy during surgery. |
---|---|
ISSN: | 1478-5951 1478-596X |
DOI: | 10.1002/rcs.1788 |