Biocompatible Multifunctional Black‐Silicon for Implantable Intraocular Sensor

Multifunctional black‐silicon (b‐Si) integrated on the surface of an implantable intraocular pressure sensor significantly improves sensor performance and reliability in six‐month in vivo studies. The antireflective properties of b‐Si triples the signal‐to‐noise ratio and increases the optical reado...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Advanced healthcare materials 2017-02, Vol.6 (4), p.np-n/a
Hauptverfasser: Lee, Jeong Oen, Narasimhan, Vinayak, Du, Juan, Ndjamen, Blaise, Sretavan, David, Choo, Hyuck
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Multifunctional black‐silicon (b‐Si) integrated on the surface of an implantable intraocular pressure sensor significantly improves sensor performance and reliability in six‐month in vivo studies. The antireflective properties of b‐Si triples the signal‐to‐noise ratio and increases the optical readout distance to a clinically viable 12 cm. Tissue growth and inflammation response on the sensor is suppressed demonstrating desirable anti‐biofouling properties.