Optical See-Through Cancer Vision Goggles Enable Direct Patient Visualization and Real-Time Fluorescence-Guided Oncologic Surgery

Background The inability to visualize the patient and surgical site directly, limits the use of current near infrared fluorescence-guided surgery systems for real-time sentinel lymph node biopsy and tumor margin assessment. Methods We evaluated an optical see-through goggle augmented imaging and nav...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Annals of surgical oncology 2017-07, Vol.24 (7), p.1897-1903
Hauptverfasser: Mondal, Suman B., Gao, Shengkui, Zhu, Nan, Habimana-Griffin, LeMoyne, Akers, Walter J., Liang, Rongguang, Gruev, Viktor, Margenthaler, Julie, Achilefu, Samuel
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Background The inability to visualize the patient and surgical site directly, limits the use of current near infrared fluorescence-guided surgery systems for real-time sentinel lymph node biopsy and tumor margin assessment. Methods We evaluated an optical see-through goggle augmented imaging and navigation system (GAINS) for near-infrared, fluorescence-guided surgery. Tumor-bearing mice injected with a near infrared cancer-targeting agent underwent fluorescence-guided, tumor resection. Female Yorkshire pigs received hind leg intradermal indocyanine green injection and underwent fluorescence-guided, popliteal lymph node resection. Four breast cancer patients received 99m Tc-sulfur colloid and indocyanine green retroareolarly before undergoing sentinel lymph node biopsy using radioactive tracking and fluorescence imaging. Three other breast cancer patients received indocyanine green retroareolarly before undergoing standard-of-care partial mastectomy, followed by fluorescence imaging of resected tumor and tumor cavity for margin assessment. Results Using near-infrared fluorescence from the dyes, the optical see-through GAINS accurately identified all mouse tumors, pig lymphatics, and four pig popliteal lymph nodes with high signal-to-background ratio. In 4 human breast cancer patients, 11 sentinel lymph nodes were identified with a detection sensitivity of 86.67 ± 0.27% for radioactive tracking and 100% for GAINS. Tumor margin status was accurately predicted by GAINS in all three patients, including clear margins in patients 1 and 2 and positive margins in patient 3 as confirmed by paraffin-embedded section histopathology. Conclusions The optical see-through GAINS prototype enhances near infrared fluorescence-guided surgery for sentinel lymph node biopsy and tumor margin assessment in breast cancer patients without disrupting the surgical workflow in the operating room.
ISSN:1068-9265
1534-4681
DOI:10.1245/s10434-017-5804-8