Axial-Plane Biomechanical Evaluation of 2 Suspensory Cortical Button Fixation Constructs for Acromioclavicular Joint Reconstruction

Background: Although numerous suture-button fixation techniques for acromioclavicular (AC) joint reconstruction have been validated with biomechanical testing in the superior direction, clinical reports continue to demonstrate high rates of construct slippage and breakage. Purpose: To compare the st...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Orthopaedic journal of sports medicine 2016-12, Vol.4 (12), p.2325967116674668-2325967116674668
Hauptverfasser: Struhl, Steven, Wolfson, Theodore S., Kummer, Frederick
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Background: Although numerous suture-button fixation techniques for acromioclavicular (AC) joint reconstruction have been validated with biomechanical testing in the superior direction, clinical reports continue to demonstrate high rates of construct slippage and breakage. Purpose: To compare the stability of a novel closed-loop double Endobutton construct with a commercially available cortical button system in both the axial and superior directions. Study Design: Controlled laboratory study. Methods: Six matched pairs of fresh-frozen cadaveric upper extremities were anatomically dissected and prepared to simulate a complete AC joint dislocation. One side of each pair was reconstructed with the double Endobutton (DE) construct and other side with the dog bone button (DB) construct. The specimens were then tested using a materials testing machine, determining initial superior and axial displacements with a preload, and then cyclically loaded in the axial direction with 70 N for 5000 cycles. Displacement was again measured with the same preloads at fixed cycle intervals. The specimens were then loaded superiorly to failure. Results: At 5000 cycles, the mean axial displacement was 1.7 mm for the DB group and 1.2 mm for the DE group (P = .19), and the mean superior displacement was 1.1 mm for the DB group and 0.7 mm for the DE group (P = .32). Load at failure was similar (558 N for DE, 552 N for DB; P = .96). There was no statistically significant difference in the modes of failure. Conclusion: Biomechanical testing of both constructs showed similar fixation stability after cyclical axial loading and similar loads to failure. Clinical Relevance: The strength of both constructs after cyclical loading in the axial plane and load-to-failure testing in the superior plane validate their continued clinical use for achieving stability in AC joint reconstruction procedures.
ISSN:2325-9671
2325-9671
DOI:10.1177/2325967116674668