A novel polymer-free ciglitazone-coated vascular stent: in vivo and ex vivo analysis of stent endothelialization in a rabbit iliac artery model
Peroxisome proliferator-activated receptor-gamma (PPARg) agonists have known pleiotropic cardiovascular effects with favourable properties in vascular remodeling, and specifically in suppression of vascular smooth muscle cell proliferation. A novel vascular stent coating using the PPARg ligand cigli...
Gespeichert in:
Veröffentlicht in: | Oncotarget 2016-09, Vol.7 (36), p.57571-57580 |
---|---|
Hauptverfasser: | , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Peroxisome proliferator-activated receptor-gamma (PPARg) agonists have known pleiotropic cardiovascular effects with favourable properties in vascular remodeling, and specifically in suppression of vascular smooth muscle cell proliferation. A novel vascular stent coating using the PPARg ligand ciglitazone (CCS) was investigated regarding its effects on endothelialization after 7 and 28 days.
Microporous bare metal stents (BMS) were coated with ciglitazone by ultrasonic flux with a load of 255 μg ciglitazone/stent. SixteenNew Zealand white rabbits, fed a with high cholesterol diet, underwent stent implantation in both iliac arteries. Everolimus-eluting stents (EES) and BMS were comparators. Histology (CD 31 immunostaining, confocal and scanning electron microscopy, morphometry) was performed after 7 and 28 days and by OCT (optical coherence tomography) in vivo after 28 days.
Microscopy showed comparable results with near complete endothelialization in CCS and BMS (%CD31 above stent struts after 7 days: 67.92±36.35 vs. 84.48±23.86; p = 0.55; endothel % above stent struts: 77.22±27.9 vs. 83.89±27.91; p = 0.78). EES were less endothelialized with minimal fibrin deposition, not found in BMS and CCS (% CD 31 above struts after 28 days, BMS: 100.0±0.0 vs. EES: 95.9±3.57 vs. CCS: 100.0±0.0; p = 0.0292). OCT revealed no uncovered struts in all stents after 28 days.
Polymer-free coating with ciglitazone, a PPARg agonist is feasible and stable over time. Our data prove unimpaired endothelial coverage of a ciglitazone-coated vascular stent system by histology and OCT. Thus, this PPARg agonist coating deserves further investigation to evaluate its potency on local neointimal suppression. |
---|---|