High-speed and high-sensitivity parallel spectral-domain optical coherence tomography using a supercontinuum light source

The three most important metrics in optical coherence tomography (OCT) are resolution, speed, and sensitivity. Because there is a complex interplay between these metrics, no previous work has obtained the best performance in all three metrics simultaneously. We demonstrate that a high-power supercon...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Optics letters 2016-12, Vol.41 (24), p.5620-5623
Hauptverfasser: Barrick, Jessica, Doblas, Ana, Gardner, Michael R, Sears, Patrick R, Ostrowski, Lawrence E, Oldenburg, Amy L
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The three most important metrics in optical coherence tomography (OCT) are resolution, speed, and sensitivity. Because there is a complex interplay between these metrics, no previous work has obtained the best performance in all three metrics simultaneously. We demonstrate that a high-power supercontinuum source, in combination with parallel spectral-domain OCT, achieves an unparalleled combination of resolution, speed, and sensitivity. This system captures cross-sectional images spanning 4  mm×0.5  mm at 1,024,000 lines/s with 2×14  μm resolution (axial×transverse) at a sensitivity of 113 dB. Imaging using the proposed system is demonstrated on highly differentiated human bronchial epithelial cells to capture and spatially localize ciliary dynamics.
ISSN:0146-9592
1539-4794
DOI:10.1364/OL.41.005620