In vitro hematocrit measurement using spectrally encoded flow cytometry

Measuring key physiological parameters of small blood samples extracted from patients could be useful for real-time clinical diagnosis at the point of care. An important parameter required from all blood tests is the blood hematocrit, a measure of the fractional volume occupied by the red cells with...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Biomedical optics express 2016-10, Vol.7 (10), p.4327-4334
Hauptverfasser: Zeidan, Adel, Golan, Lior, Yelin, Dvir
Format: Artikel
Sprache:eng
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Measuring key physiological parameters of small blood samples extracted from patients could be useful for real-time clinical diagnosis at the point of care. An important parameter required from all blood tests is the blood hematocrit, a measure of the fractional volume occupied by the red cells within the blood. In this work, we present a method for evaluation of hematocrit based on the data acquired using spectrally encoded flow cytometry. Analysis of the reflectance confocal images of blood within a flow chamber resulted in an error as low as 1.7% in the measured hematocrit. The technique could be used as part of an diagnostic system that measures important blood parameters at the point of care.
ISSN:2156-7085
2156-7085
DOI:10.1364/BOE.7.004327