Continuous venovenous hemofiltration using customized replacement fluid for acute kidney injury with severe hypernatremia

The initiation of continuous renal replacement therapy (CRRT) in acute kidney injury (AKI) with severe hypernatremia is challenging since sodium concentrations in commercial replacement fluid (RF) and dialysate solutions are usually fixed at 140 mEq/L. We present a case of AKI with severe hypernatre...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Clinical kidney journal 2016-08, Vol.9 (4), p.540-542
Hauptverfasser: Paquette, François, Goupil, Rémi, Madore, François, Troyanov, Stéphan, Bouchard, Josée
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The initiation of continuous renal replacement therapy (CRRT) in acute kidney injury (AKI) with severe hypernatremia is challenging since sodium concentrations in commercial replacement fluid (RF) and dialysate solutions are usually fixed at 140 mEq/L. We present a case of AKI with severe hypernatremia successfully treated with CRRT using commercial RF solutions customized to prevent rapid correction of hypernatremia. None of the few case reports published on hypernatremia and AKI requiring CRRT have included formulas to help modulate the sodium content in the solutions. We present an equation to facilitate adjustment of the sodium concentration in this setting.
ISSN:2048-8505
2048-8513
DOI:10.1093/ckj/sfw036