A comprehensive assay for CFTR mutational analysis using next-generation sequencing
Cystic fibrosis is a life-threatening genetic disorder that has been associated with mutations in the CFTR [cystic fibrosis transmembrane conductance regulator (ATP-binding cassette sub-family C, member 7)] gene. Hundreds of CFTR mutations have been detected to date. Current CFTR genotyping assays t...
Gespeichert in:
Veröffentlicht in: | Clinical chemistry (Baltimore, Md.) Md.), 2013-10, Vol.59 (10), p.1481-1488 |
---|---|
Hauptverfasser: | , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Cystic fibrosis is a life-threatening genetic disorder that has been associated with mutations in the CFTR [cystic fibrosis transmembrane conductance regulator (ATP-binding cassette sub-family C, member 7)] gene. Hundreds of CFTR mutations have been detected to date. Current CFTR genotyping assays target a subset of these mutations, particularly a mutation panel recommended by the American College of Medical Genetics for carrier screening of the general population. Fast sequencing of the entire coding sequence in a scalable manner could expand the detection of CFTR mutations and facilitate management of costs and turnaround times in the clinical laboratory.
We describe a proof-of-concept CFTR assay that uses PCR target enrichment and next-generation sequencing on the Ion Torrent Personal Genome Machine™ (PGM™) platform.
The scalability of the assay was demonstrated, with an average mean depth of coverage ranging from 500× to 3500×, depending on the number of multiplexed patient samples and the Ion Torrent chip used. In a blinded study of 79 previously genotyped patient DNA samples and cell lines, our assay detected most of the mutations, including single-nucleotide variants, small insertions and deletions, and large copy-number variants. The reproducibility was 100% for detecting mutations in independent runs. Our assay demonstrated high specificity, with only 2 false-positive calls (at 2184delA) found in 2 samples caused by a sequencing error in a homopolymer stretch of sequence. The detection rate for variants of unknown significance was very low in the targeted region.
With continued optimization and system refinements, PGM sequencing promises to be a powerful, rapid, and scalable means of clinical diagnostic sequencing. |
---|---|
ISSN: | 0009-9147 1530-8561 |
DOI: | 10.1373/clinchem.2013.206466 |