Optical-Fiber Power Meter Comparison between NIST and KRISS
We describe the results of a comparison of reference standards between the National Institute of Standards and Technology (NIST-USA) and Korea Research Institute of Standards and Science (KRISS-R.O. Korea) for optical fiber-based power measurements at wavelengths of 1302 nm and 1546 nm. We compare t...
Gespeichert in:
Veröffentlicht in: | Journal of research of the National Institute of Standards and Technology 2012-11, Vol.117 (6), p.328-332 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We describe the results of a comparison of reference standards between the National Institute of Standards and Technology (NIST-USA) and Korea Research Institute of Standards and Science (KRISS-R.O. Korea) for optical fiber-based power measurements at wavelengths of 1302 nm and 1546 nm. We compare the laboratories' reference standards by means of a temperature-controlled optical trap detector. Measurement results showed the largest difference of less than 2.5 parts in 10(3), which is within the combined standard (k=1) uncertainty for the two laboratories' reference standards. |
---|---|
ISSN: | 1044-677X 2165-7254 2165-7254 |
DOI: | 10.6028/jres.117.019 |