The Minimum Duration of Sensor Data From Which Glycemic Variability Can Be Consistently Assessed
Background: Despite much discussion regarding the clinical relevance of glycemic variation (GV), little discourse has addressed the properties of the data set from which it is derived. We aimed to assess the minimum duration of data required using continuous glucose monitoring (CGM) that most closel...
Gespeichert in:
Veröffentlicht in: | Journal of diabetes science and technology 2014-03, Vol.8 (2), p.273-276 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Background:
Despite much discussion regarding the clinical relevance of glycemic variation (GV), little discourse has addressed the properties of the data set from which it is derived. We aimed to assess the minimum duration of data required using continuous glucose monitoring (CGM) that most closely approximates to a gold standard 90-day measure.
Methods:
Data from 20 children and adolescents with type 1 diabetes were examined. All participants had CGM data sets of 90 days duration, from which standard deviation (SD), coefficient of variation (CV), mean amplitude of glycemic action (MAGE), and continuous overlapping net glycemic action (CONGA1-8) were calculated for the overall period and then investigational periods of 2, 4, 6, 12, 18, 24, and 30 days. The percentage difference between each measure and the overall measure per time period was assessed.
Results:
As the duration of the CGM data set increased, the percentage error continued to decrease, giving a metric approximating more closely toward the overall measure. Median SD and CV differed from the overall measure by |
---|---|
ISSN: | 1932-2968 1932-2968 1932-3107 |