Biomechanical Evaluation of Metacarpal Fracture Fixation: Application of a 90° Internal Fixation Model

Purpose Complications in metacarpal fracture treatment increase in proportion to the severity of the initial injury and the invasiveness of the surgical fixation technique. This manuscript evaluates the feasibility of minimizing internal fixation construct size and soft tissue dissection, while pres...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Hand (New York, N.Y.) N.Y.), 2015-03, Vol.10 (1), p.94-99
Hauptverfasser: Watt, Andrew J., Ching, Randal P., Huang, Jerry I.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Purpose Complications in metacarpal fracture treatment increase in proportion to the severity of the initial injury and the invasiveness of the surgical fixation technique. This manuscript evaluates the feasibility of minimizing internal fixation construct size and soft tissue dissection, while preserving the advantages of stable internal fixation in a biomechanical model. We hypothesized that comparable construct stability could be achieved with mini-plates in an orthogonal (90/90) configuration compared with a standard dorsal plating technique. Methods This hypothesis was evaluated in a transverse metacarpal fracture model. Twelve metacarpals were subject to either placement of a 2.0-mm six-hole dorsal plate or two 1.5-mm four-hole mini-plates in a 90/90 configuration. These constructs were tested to failure in a three-point bending apparatus, attaining failure force, displacement, and stiffness. Results Mean failure force was 353.5±121.1 N for the dorsal plating construct and 358.8±77.1 N for the orthogonal construct. Mean failure displacement was 3.3±1.2 mm for the dorsal plating construct and 4.1±0.9 mm for the orthogonal construct. Mean stiffness was 161.3±50.0 N/mm for the dorsal plating construct and 122.1±46.6 N/mm for the orthogonal construct. Mean failure moment was 3.09±1.06 Nm for the dorsal plating construct and 3.14±0.67 Nm for the orthogonal construct. The dorsal plating group failed via screw pullout, whereas the orthogonal failed either by screw pullout or breakage of the plate. Conclusions When subject to apex dorsal bending, the orthogonal construct and the standard dorsal plate construct behaved comparably. These data suggest that despite its shorter length, lower profile, and less substantial screws, the orthogonal construct provides sufficient rigidity. Clinical Relevance This study represents a “proof of concept” regarding the applicability of orthogonal plating in the metacarpal and provides the foundation for minimizing construct size and profile.
ISSN:1558-9447
1558-9455
DOI:10.1007/s11552-014-9673-3