Reliability of using a fixed matrix in coregistration of combined PET–MRI in a split magnet design
We consider the effects of using a fixed linear transformation to match positron emission tomography (PET) and magnetic resonance imaging (MRI) data acquired simultaneously using a split-magnet system. Estimates of the frequency offset in MRI scans were used to calculate geometric variability in MRI...
Gespeichert in:
Veröffentlicht in: | Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment Accelerators, spectrometers, detectors and associated equipment, 2013-02, Vol.702, p.54-55 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We consider the effects of using a fixed linear transformation to match positron emission tomography (PET) and magnetic resonance imaging (MRI) data acquired simultaneously using a split-magnet system. Estimates of the frequency offset in MRI scans were used to calculate geometric variability in MRI reconstruction as a consequence of mis-setting this parameter in addition to repeated estimation of the transformation matrix by manual measurements. None of the measured variability approached the resolution of the PET images, so we concluded that a fixed matrix can be reliably used in such a system. |
---|---|