Ventricular perforation by pacemaker lead repaired with two hemostatic devices
Abstract INTRODUCTION Cardiac perforation is a rare, but potentially serious, complication of pacemaker implantation that may develop days or weeks after implantation. PRESENTATION OF CASE In the current case, 92-year-old man underwent permanent pacemaker implantation, but he presented 3 weeks later...
Gespeichert in:
Veröffentlicht in: | International journal of surgery case reports 2014-01, Vol.5 (12), p.906-908 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Abstract INTRODUCTION Cardiac perforation is a rare, but potentially serious, complication of pacemaker implantation that may develop days or weeks after implantation. PRESENTATION OF CASE In the current case, 92-year-old man underwent permanent pacemaker implantation, but he presented 3 weeks later with severe symptoms. Computed tomography showed protrusion of the tip of the ventricular electrode through the right ventricle and into the chest wall. During an urgent surgical intervention, the lead was disconnected and extracted. A sealing hemostatic device and an hemostatic patch were applied to repair the ventricle; the procedure was uneventfull. DISCUSSION This case demonstrates how the correct diagnosis of ventricular perforation is crucial, and should be followed immediately by surgical planning. CONCLUSION The hemostatic patch is a valuable alternative to sutures in patients with thin and fragile ventricular wall, unable to undergo stitching. |
---|---|