Improvement of therapeutic index for brain tumors with daily image guidance

Image-guidance maximizes the therapeutic index of brain irradiation by decreasing setup uncertainty. As dose-volume data emerge defining the tolerance of critical normal structures responsible for neuroendocrine function and neurocognition, minimizing clinical target volume (CTV) to planning target...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Radiation oncology (London, England) England), 2013-12, Vol.8 (1), p.283-283, Article 283
Hauptverfasser: Shields, Lisa Be, Coons, James M, Dedich, Catherine, Ragains, Maria, Scalf, Kristi, Vitaz, Todd W, Spalding, Aaron C
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Image-guidance maximizes the therapeutic index of brain irradiation by decreasing setup uncertainty. As dose-volume data emerge defining the tolerance of critical normal structures responsible for neuroendocrine function and neurocognition, minimizing clinical target volume (CTV) to planning target volume (PTV) expansion of targets near these structures potentially lessens long-term toxicity. We reviewed the treatment records of 29 patients with brain tumors, with a total of 517 fractions analyzed. The CTV was uniformly expanded by 3 mm to create the PTV for all cases. We determined the effect of patient specific factors (prescribed medications, weight gain, tumor location) and image-guidance technique on setup uncertainty and plotted the mean +/- standard deviation for each factor. ANOVA was used to determine significance between these factors on setup uncertainty. We determined the impact of applying the initial three fraction variation as custom PTV-expansion on dose to normal structures. The initial 3 mm margin encompassed 88% of all measured shifts from daily imaging for all fractions. There was no difference (p = n.s.) in average setup uncertainty between CBCT or kV imaging for all patients. Vertical, lateral, longitudinal, and 3D shifts were similar (p = n.s.) between days 1, 2, and 3 imaging and later fractions. Patients prescribed sedatives experienced increased setup uncertainty (p 
ISSN:1748-717X
1748-717X
DOI:10.1186/1748-717X-8-283