Evaluation of Stability of Rotating Hinge Knee Prostheses: A Biomechanical Study
Purpose. Rotating hinge knee prostheses should provide a stable situation following reconstruction. We performed a biomechanical analysis to establish the association between design of the central rotational stem (peg) and implant’s stability, in a theoretical setting. Methods. Six different rotatin...
Gespeichert in:
Veröffentlicht in: | ISRN orthopedics 2013-07, Vol.2013, p.701693-7 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Purpose. Rotating hinge knee prostheses should provide a stable situation following reconstruction. We performed a biomechanical analysis to establish the association between design of the central rotational stem (peg) and implant’s stability, in a theoretical setting. Methods. Six different rotating hinge designs were tested, and three observers performed two different measurements with a custom made biomechanical apparatus and laterally directed pressure. The aim was to assign the degree of tilting of the peg within the vertical post-in channel by extending the distraction as well as the maximum amount of distraction before the peg’s dislocation. An intraclass-correlation coefficient (ICC) was calculated to determine the observer’s reliability. Results. Implant designs with cylindrical pegs of different lengths were superior to implant designs with conical or other shaped pegs concerning stability and maximum amount of distraction before dislocation, showing steep rising distraction-angular displacement curves. The ICC at 15 mm and 25 mm of distraction revealed high interobserver reliability (P |
---|---|
ISSN: | 2090-6161 2090-617X 2090-617X |
DOI: | 10.1155/2013/701693 |