Validation of a portable nitric oxide analyzer for screening in primary ciliary dyskinesias
Nasal nitric oxide (nNO) levels are very low in primary ciliary dyskinesia (PCD) and it is used as a screening test. We assessed the reliability and usability of a hand-held analyser in comparison to a stationary nitric oxide (NO) analyser in 50 participants (15 healthy, 13 PCD, 22 other respiratory...
Gespeichert in:
Veröffentlicht in: | BMC pulmonary medicine 2014-02, Vol.14 (1), p.18-18, Article 18 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Nasal nitric oxide (nNO) levels are very low in primary ciliary dyskinesia (PCD) and it is used as a screening test.
We assessed the reliability and usability of a hand-held analyser in comparison to a stationary nitric oxide (NO) analyser in 50 participants (15 healthy, 13 PCD, 22 other respiratory diseases; age 6-79 years). Nasal NO was measured using a stationary NO analyser during a breath-holding maneuver, and using a hand-held analyser during tidal breathing, sampling at 2 ml/sec or 5 ml/sec. The three methods were compared for their specificity and sensitivity as a screen for PCD, their success rate in different age groups, within subject repeatability and acceptability. Correlation between methods was assessed.
Valid nNO measurements were obtained in 94% of participants using the stationary analyser, 96% using the hand-held analyser at 5 ml/sec and 76% at 2 ml/sec. The hand-held device at 5 ml/sec had excellent sensitivity and specificity as a screening test for PCD during tidal breathing (cut-off of 30 nL/min,100% sensitivity, >95% specificity). The cut-off using the stationary analyser during breath-hold was 38 nL/min (100% sensitivity, 95% specificity). The stationary and hand-held analyser (5 ml/sec) showed reasonable within-subject repeatability(% coefficient of variation = 15).
The hand-held NO analyser provides a promising screening tool for PCD. |
---|---|