Structural Analysis for Wingspan Stent in a Perforator Model

Perforator infarction represents a critical problem after intracranial Wingspan stent. To explore the mechanism of perforator infarction, we simulated the stent-artery interaction at an atheromatous plaque with perforator. Structural deformation and biomechanical stress distribution after stenting w...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Interventional neuroradiology 2013-09, Vol.19 (3), p.271-275
Hauptverfasser: Fujimoto, Motoaki, Shobayashi, Yasuhiro, Takemoto, Koichiro, Tateshima, Satoshi, Viñuela, Fernando
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Perforator infarction represents a critical problem after intracranial Wingspan stent. To explore the mechanism of perforator infarction, we simulated the stent-artery interaction at an atheromatous plaque with perforator. Structural deformation and biomechanical stress distribution after stenting were analyzed. High radial stress values were located along the stent struts, which surrounded the area with high circumferential stress. Stretched perforator orifice in a circumferential direction after stenting was simulated. These results show that structural deformation could play a role in the mechanism of perforator occlusion after Wingspan stenting.
ISSN:1591-0199
2385-2011
DOI:10.1177/159101991301900302