Adiabatic inversion pulses for myocardial T1 mapping
Purpose To evaluate the error in T1 estimates using inversion‐recovery‐based T1 mapping due to imperfect inversion and to perform a systematic study of adiabatic inversion pulse designs in order to maximize inversion efficiency for values of transverse relaxation (T2) in the myocardium subject to a...
Gespeichert in:
Veröffentlicht in: | Magnetic resonance in medicine 2014-04, Vol.71 (4), p.1428-1434 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Purpose
To evaluate the error in T1 estimates using inversion‐recovery‐based T1 mapping due to imperfect inversion and to perform a systematic study of adiabatic inversion pulse designs in order to maximize inversion efficiency for values of transverse relaxation (T2) in the myocardium subject to a peak power constraint.
Methods
The inversion factor for hyperbolic secant and tangent/hyperbolic tangent adiabatic full passage waveforms was calculated using Bloch equations. A brute‐force search was conducted for design parameters: pulse duration, frequency range, shape parameters, and peak amplitude. A design was selected that maximized the inversion factor over a specified range of amplitude and off‐resonance and validated using phantom measurements. Empirical correction for imperfect inversion was performed.
Results
The tangent/hyperbolic tangent adiabatic pulse was found to outperform hyperbolic secant designs and achieve an inversion factor of 0.96 within ±150 Hz over 25% amplitude range with 14.7 µT peak amplitude. T1 mapping errors of the selected design due to imperfect inversion was ∼4% and could be corrected to |
---|---|
ISSN: | 0740-3194 1522-2594 |
DOI: | 10.1002/mrm.24793 |