Accuracy and reliability of continuous glucose monitoring systems: a head-to-head comparison
This study assessed the accuracy and reliability of three continuous glucose monitoring (CGM) systems. We studied the Animas® (West Chester, PA) Vibe™ with Dexcom® (San Diego, CA) G4™ version A sensor (G4A), the Abbott Diabetes Care (Alameda, CA) Freestyle® Navigator I (NAV), and the Medtronic (Nort...
Gespeichert in:
Veröffentlicht in: | Diabetes technology & therapeutics 2013-08, Vol.15 (8), p.722-726 |
---|---|
Hauptverfasser: | , , , , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | This study assessed the accuracy and reliability of three continuous glucose monitoring (CGM) systems.
We studied the Animas® (West Chester, PA) Vibe™ with Dexcom® (San Diego, CA) G4™ version A sensor (G4A), the Abbott Diabetes Care (Alameda, CA) Freestyle® Navigator I (NAV), and the Medtronic (Northridge, CA) Paradigm® with Enlite™ sensor (ENL) in 20 patients with type 1 diabetes mellitus. All systems were investigated both in a clinical research center (CRC) and at home. In the CRC, patients received a meal with a delayed and increased insulin dose to induce a postprandial glucose peak and nadir. Hereafter, randomization determined which two of the three systems would be worn at home until the end of functioning, attempting use beyond manufacturer-specified lifetime. Patients performed at least five reference finger sticks per day. An analysis of variance was performed on all data points ≥15 min apart.
Overall average mean absolute relative difference (MARD) (SD) measured at the CRC was 16.5% (14.3%) for NAV and 16.4% (15.6%) for ENL, outperforming G4A at 20.5% (18.2%) (P |
---|---|