High-speed, clinical-scale microfluidic generation of stable phase-change droplets for gas embolotherapy
In this study we report on a microfluidic device and droplet formation regime capable of generating clinical-scale quantities of droplet emulsions suitable in size and functionality for in vivo therapeutics. By increasing the capillary number-based on the flow rate of the continuous outer phase-in o...
Gespeichert in:
Veröffentlicht in: | Lab on a chip 2011-12, Vol.11 (23), p.399-3998 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | In this study we report on a microfluidic device and droplet formation regime capable of generating clinical-scale quantities of droplet emulsions suitable in size and functionality for
in vivo
therapeutics. By increasing the capillary number-based on the flow rate of the continuous outer phase-in our flow-focusing device, we examine three modes of droplet breakup: geometry-controlled, dripping, and jetting. Operation of our device in the dripping regime results in the generation of highly monodisperse liquid perfluoropentane droplets in the appropriate 3-6 μm range at rates exceeding 10
5
droplets per second. Based on experimental results relating droplet diameter and the ratio of the continuous and dispersed phase flow rates, we derive a power series equation, valid in the dripping regime, to predict droplet size,
D
d
≅ 27(
Q
C
/
Q
D
)
−5/12
. The volatile droplets in this study are stable for weeks at room temperature yet undergo rapid liquid-to-gas phase transition, and volume expansion, above a uniform thermal activation threshold. The opportunity exists to potentiate locoregional cancer therapies such as thermal ablation and percutaneous ethanol injection using thermal or acoustic vaporization of these monodisperse phase-change droplets to intentionally occlude the vessels of a cancer.
We report on a microfluidic device and droplet formation regime capable of generating clinical-scale quantities of liquid perfluorocarbon phase-change droplets suitable in size and functionality for
in vivo
gas embolotherapy. |
---|---|