320-detector row CT coronary angiography: effects of heart rate and heart rate variability on image quality, diagnostic accuracy and radiation exposure

To evaluate the effects of heart rate and heart rate variability on image quality, patient dose and diagnostic accuracy of 320-detector row CT. 94 patients were prospectively enrolled. Heart rate was defined as the mean value of different intervals elapsing between two consecutive R waves in an elec...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:British journal of radiology 2012-08, Vol.85 (1016), p.e388-e394
Hauptverfasser: Sun, G, Li, M, Jiang, X-S, Li, L, Peng, Z-H, Li, G-Y, Xu, L
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:To evaluate the effects of heart rate and heart rate variability on image quality, patient dose and diagnostic accuracy of 320-detector row CT. 94 patients were prospectively enrolled. Heart rate was defined as the mean value of different intervals elapsing between two consecutive R waves in an electrocardiogram (R-R intervals) and the heart rate variability was calculated as the standard deviation from the average heart rate. The image quality was evaluated by four grades, according to motion artefacts ("step artefacts" and "blurring artefacts"). The diagnostic accuracy was analysed in 43 patients who were scheduled for invasive coronary angiography (ICA). The coeffects of heart rate and heart rate variability on image quality, radiation dose and diagnostic accuracy were evaluated by multivariate regression. The mean image quality score was 1.2 ± 0.5 and the mean effective dose was 14.8 ± 9.8 mSv. The results showed that heart rate (74.0 ± 11.2 beats per minute) was the single factor influencing image quality (p
ISSN:0007-1285
1748-880X
DOI:10.1259/bjr/92160185