A distal femoral supra-condylar plate: biomechanical comparison with condylar plate and first clinical application for treatment of supracondylar fracture

Purpose An anatomical supra-condylar plate is designed and analysed by biomechanical testing. Methods The biomechanical properties of the supra-condylar and condylar plate were compared in six matched pairs of cadaveric femurs. A transverse osteotomy gap was created to simulate an OTA/AO type A3 sup...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:International orthopaedics 2012-08, Vol.36 (8), p.1673-1679
Hauptverfasser: Liang, Bowei, Ding, Zhenqi, Shen, Junguo, Zhai, Wenliang, Kang, Liangqi, Zhou, Liang, Sha, Mo, Liang, Dongzhu
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Purpose An anatomical supra-condylar plate is designed and analysed by biomechanical testing. Methods The biomechanical properties of the supra-condylar and condylar plate were compared in six matched pairs of cadaveric femurs. A transverse osteotomy gap was created to simulate an OTA/AO type A3 supracondylar fracture. The left and right specimens were fitted with supra-condylar and condylar plate, respectively. Nondestructive axial compression, three-point bending and torsion tests were performed, and the peak load of the bone-implant construction was measured. The fracture site suitable for supra-condylar plate application and its correlation with femoral length were calculated. The gender influence on it was also discussed. Results The difference of stiffness between the supra-condylar and condyle groups were not significant ( P  > 0.05) at 363.4 and 362.5 N/mm for compression, 229.5 and 237.6 N/mm in the sagittal plane and 195.5 and 188.4 N/mm in the coronal plane for three-point bending, and 7.5 and 7.9 Nm/deg for axial torsion, respectively. The peak load was 4438 ± 136.15 N and 5215 ± 174.33 N, respectively, for the two groups. The average extent of the fracture site suitable for the application of the supra-condylar plate was 70.86 ± 4.61 mm. The femoral length and gender showed no influence on it. Conclusion Despite the limited bone contact area provided by the supra-condylar plate, its construct stiffness is comparable to the condylar plate. The supra-condylar plate can be used to treat carefully-selected extra-articular supracondylar fractures.
ISSN:0341-2695
1432-5195