Electrochemical corrosion assessment of RaCe and Mtwo rotary nickle-titanium instruments after clinical use and sterilization
The aim of the present study was to electrochemically evaluate corrosion resistance of RaCe and Mtwo files after repeated sterilization and preparation procedures. A total of 450 rotary files were used. In the working groups, 72 files from each file type were distributed into 4 groups. RaCe and Mtwo...
Gespeichert in:
Veröffentlicht in: | Medicina oral, patología oral y cirugía bucal patología oral y cirugía bucal, 2012-03, Vol.17 (2), p.e331-e336 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The aim of the present study was to electrochemically evaluate corrosion resistance of RaCe and Mtwo files after repeated sterilization and preparation procedures.
A total of 450 rotary files were used. In the working groups, 72 files from each file type were distributed into 4 groups. RaCe and Mtwo files were used to prepare one root canal of the mesial root of extracted human mandibular first molars. The procedure was repeated to prepare 2 to 8 canals. The following irrigation solutions were used: group 1, RaCe files with 2.5% NaOCl; group 2, RaCe files with normal saline; group 3, Mtwo files with 2.5% NaOCl; and group 4, Mtwo files with normal saline in the manner described. In autoclave groups, 72 files from each file type were evenly distributed into 2 groups. Files were used for a cycle of sterilization without the use of files for root canal preparation. Nine new unused files from each file type were used as controls. Then the instruments were sent for corrosion assessment. Mann-Whitney U and Wilcoxon tests were used for independent and dependent groups, respectively.
Statistical analysis indicated that there were significant differences in corrosion resistance of files associated with working and autoclave groups between RaCe and Mtwo file types (p |
---|---|
ISSN: | 1698-6946 1698-4447 1698-6946 |
DOI: | 10.4317/medoral.17413 |